EasyManua.ls Logo

Dräger Atlan A300 - Page 141

Dräger Atlan A300
332 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 141
Tests
8.4.3 Breathing circuit
Prerequisite:
Breathing system is complete and locked.
Breathing system cover is fitted.
1. Hoses (1) and filters, e.g., at the Y-piece (2), are connected properly.
2. Extendable hoses (3) are extended to the length intended for use.
Do not change the length of the hoses after the test is done.
3. Remove the water from the hoses (4).
4. Prerequisite:
Patient-gas measurement module present
Setting: Sample line is connected during test: Off
Setting: Test breathing circuit: On
Electronically controlled gas mixer or mechanically controlled gas mixer with
O
2 detection
3511535117
2
1
3
4

Table of Contents

Other manuals for Dräger Atlan A300

Related product manuals