Quick reference guide | Atlan SW 2.0n 17
Assembly and preparation
When taking a gas sample at a connector
located as close as possible to the patient
One breathing system filter on the inspi-
ratory port and one on the expiratory port
Use only heat and moisture exchangers
(HMEs) between the Y-piece and the
patient that are listed in the list of acces-
sories.
Do not use fine-pored filters or HME fil-
ters between the Y-piece and the patient.
Set the alarm limits for MV low and Paw
to suitable values. Observe the following
information: "Optimization of the sample
gas quality", page 40.
XX
When using fine-pored filters between the Y-
piece and the patient
Do not connect the sample line between
the tube and the filter. Instead, connect it
to the filter or the Y-piece.
XX
Configuration Adult Ped. Neo.