50 Technical manual | X-am® 3500/8000
Maintenance
6. Rotate the gas detector and carefully lift it out of the upper cradle.
7. Exchanging the EC sensors:
a. Remove the sensor sealing plate from the EC sensors.
b. Place the special tool on the trough. Using the accompanying special tool (1),
carefully lift out the desired EC sensor (2). Do not bend the sensor pins.
c. Only if necessary: Exchange the white sensor holder.
d. Have the sensor type and code (marked on the new sensor) as well as the
sensor position at hand for the new sensor inauguration.
3500335004
1
2

Table of Contents