60 Technical manual | X-am® 3500/8000
Maintenance
9. Insert the new pump into the existing pump block and mount the entire
assembly.
10.Re-assemble the gas detector. Make sure that the connector cable of the pump
is running in the correct direction. The connecting cables must be routed within
the space between the pump and the pump block. Observe the cable routing in
figure (2).
When assembling the options board cover, make sure that the differential
pressure sensor is positioned correctly.
6.9.10 Horn / resonator
Requirements:
The gas detector is open.
1. Loosen 3 screws (Torx T8) on the options board.
2. Rotate the gas detector and carefully lift it out of the front housing cradle.
3. Remove the 3 clips from the options board cover and remove the options board.
4. Remove the horn (1) from the options board.
5. Insert a new horn if necessary.
6. Exchange the resonator (2) if necessary.