Electrical operation time
Oxylog 2000 can be used for about six hours with a fully
charged pack of rechargeable NiCd batteries.
The electrical operation time with NiCd batteries can
decline as the battery capacity is reduced with time.
Oxylog 2000 can be used for about four hours with new
alkaline manganese batteries.
The average pneumatic operation time for ventilating adults
(minute volume MV = 10 L/min) from a 2.5 L/200 bar
cylinder is approx. 45 minutes, see page 26.
The following display appears when the electrical
operation time runs down:
or
The remaining electrical operation time equals approx.
10 minutes.
The illumination in the display cannot be switched on
during this time.
If necessary:
Continue ventilation with a self-inflating manual
ventilation bag.
Fit a new pack of alkaline manganese batteries,
page 23
or
connect to the external power supply, page 21.
15
Operation
Electrical operation time
Charge NiCd
Change bat.