Fitting/replacing internal NiCd battery pack
Before using ventilator for the first time
When the following display appears during charging:
and
Every 2 years as a precautionary measure (check in
medicament logbook).
Undo screw in cover of battery compartment, e.g.
with a coin, and remove cover.
Remove faulty NiCd battery pack and disconnect
plug.
Connect plug of new NiCd battery pack to socket of
battery compartment and place battery pack in
compartment.
Refit cover and tighten screw.
Charge built-in new battery pack:
– with DC voltage from the on-board supply
– or with the power supply unit.
Dispose of the faulty NiCd battery pack as special
waste, page 37.
Charge NiCd battery pack, page 24.
Screw cover into place.
Replacing the fuse
if operation with NiCd or alkaline manganese batteries is
impossible.
Remove defective fuse with the aid of a screwdriver.
Press new fuse into terminal.
Quick-acting fuse: F 1L 250V IEC 127
Delay-action fuse: T 1L 250V IEC 127
Screw cover into place.
38
Maintenance intervals
Fitting/replacing internal NiCd battery pack
Replacing the fuse