35
Faultt – Cause – Remedy
Fault / Message Cause Remedy
Main power down
Alarm indicator flashes,
intermittent tone sounds.
Plug connection to external power
supply does not make contact.
Power supply unit or DC/DC
converter defective.
Check plug connection or replace
power supply unit or DC/DC
converter.
Press »g Reset« to confirm.
Ventilator continues operation with
internal power supply.
NiCd discharged
Alarm indicator flashes,
intermittent tone sounds,
ventilation ceases.
NiCd battery pack flat, no external
power supply
Replace NiCd battery pack or
connect power supply.
No NiCd charge
This message only appears in
standby mode.
NiCd battery pack not fitted or
alkaline manganese battery pack
fitted or ambient temperature when
charging the internal NiCd battery
pack is outside the range 0 to 35 °C
or internal fuse defective.
Charge NiCd battery pack at ambient
temperatures between 0 and 35 °C.
No NiCd or bat. Ventilator connected to external
power supply, internal power supply
not available.
Fit fully charged NiCd or new alkaline
manganese battery pack.
Paw high
Alarm indicator flashes,
intermittent tone sounds.
The minute volume cannot
be fully applied.
Single beep
Stenosis in the airways. Ventilation
hose kinked. Reduced lung
compliance. The inspiratory flow
resulting from the setting for V
T,
Freq. and TI : TE is too high.
Patient "fights" the machine.
80 mbar set – for cardio-pulmonary
resuscitation.
Clear airways.
Ensure ventilation hose is not kinked.
Set higher Pmax value. Set a longer
inspiration time (smaller flow) with the
TI : TE knob.
Change ventilation pattern or
ventilation mode. Sedate patient if
necessary.
No remedy required.
Paw low
Alarm indicator flashes,
intermittent tone sounds.
Disconnection / leakage in patient
connection, breathing valve or
ventilation hose.
Diaphragm wrongly fitted in breathing
valve or damaged.
Breathing valve cover damaged/worn
Leak in cuff.
Ensure connections are tight.
Fit diaphragm correctly or replace it.
Replace breathing valve.
Inflate cuff and check for leaks.