Technical dataen
10.2 Technical data of the sensors and the measuring device settings for Pac 6000/6500
CO-LC H
2
S-LC O
2
SO
2
Measuring range 0 ... 1999 ppm 0...100ppm 2 ... 25 Vol% 0...100ppm
Calibration gas concentration 50 ppm 20 ppm 18 Vol% 10 ppm
Temperature range, operation -40 ... 50 °C
-40 ... 122 °F
-40 ... 50 °C
-40 ... 122 °F
-40 ... 50 °C
-40 ... 122 °F
-30 ... 50 °C
-22 ... 122 °F
Alarm threshold A1
1)
1) Observe special, customer-requested settings. Device parameters can be changed with the Dräger CC-Vision Basic PC software.
30 ppm 5 ppm 19 Vol%
2)
2) For O
2
, A1 is the lower alarm threshold for displaying oxygen deficiency.
1 ppm
Acknowledgeable -
Latching
-
- -
Alarm threshold A2
1)
60 ppm 10 ppm 23 Vol% 2 ppm
Acknowledgeable - - - -
Latching
TWA threshold value
1) 3)
3) Not applicable for Pac 6000.
30 ppm 10 ppm - 1 ppm
STEL threshold value
1) 3)
60 ppm 10 ppm - 1 ppm
Mean value period 15 min 15 min - 15 min
Measurement accuracy
Zero point: ±2 ppm ±0.1 ppm ±0.2 Vol% ±0.1 ppm
Sensitivity: [% of the measured value] ±2 ± 5 ± 1 ±2
Long-term drift (20 °C)
Zero point: ±2 ppm/a ±0.2 ppm/a ±0.5 Vol%/a ±1 ppm/a
Sensitivity: [% of measured value/year] ±3 ±5 ±1 ±2/month
Sensor item number
4)
4) The sensors have a limited life span. Excessive storage periods compromise the operational life time of the sensors. The adequate temperature range for storage is 0 ... 35 °C (32 ... 95 °F).
68 13 210 68 11 525 68 10 881 68 10 885
Sensor data sheet item number 90 33 454 90 23 970 90 23 820 90 23 919