Technical data
21
en
10.3.1 Sensors for Pac 8500 DUAL
H
2
S LC / CO LC O
2
/ CO-LC CO H
2
-CP
Measuring range 0 ... 100 ppm H
2
S
0 ... 2000 ppm CO
0...25Vol% O
2
0 ... 2000 ppm CO
0 ... 2000 ppm
Calibration gas concentration 5 ... 90 ppm H
2
S
20 ... 450 ppm CO
12 ... 20 Vol% O
2
20 ... 1800 ppm CO
1000 ppm H
2
20 ... 1800 ppm CO
Temperature range, operation -40 ... 50 °C
-40...122°F
-40 ... 50 °C
-40 ... 122 °F
-40 ... 50 °C
-40...122°F
Alarm threshold A1
1)
1) Observe special, customer-requested settings. Device parameters can be changed with the Dräger CC-Vision Basic PC software.
5 ppm H
2
S
30 ppm CO
19 O
2
30 CO
30 ppm
Acknowledgeable O
2
:
-
CO:
Latching - O
2
:
CO: -
-
Alarm threshold A2
1)
10 ppm H
2
S
60 ppm CO
23 O
2
60 CO
60 ppm
Acknowledgeable - O
2
:
-
CO: -
-
Latching O
2
:
CO:
TWA threshold value
1)
5 ppm H
2
S
30 ppm CO
-
30 ppm CO
-
30 ppm CO
STEL threshold value
1)
10 ppm H
2
S
60 ppm CO
-
60 ppm CO
-
60 ppm CO
Mean value period 15 min 15 min 15 min
Measurement accuracy
Zero point: H
2
S: 0.4 ppm
CO: 2ppm
O
2
±0.4 Vol%
CO: ±2 ppm
±6 ppm
Sensitivity: [% of the measured value] H
2
S: ±5
CO: ±2
O
2
±1
CO: ±2
±2
Long-term drift (20 °C)
Zero point: H
2
S: ±0.2 ppm/a
CO: ±2 ppm/a
O
2
±0.5 Vol%/a
CO: ±2 ppm/a
±2 ppm/a
Sensitivity:
[% of measured value/year
] H
2
S: ±5
CO: ±3
O
2
±1
CO: ±3
±1/month
Sensor item number
2)
2) The sensors have a limited life span. Excessive storage periods compromise the operational life time of the sensors. The adequate temperature range for storage
is 0 ... 35 °C (32 ... 95 °F).
68 13 280 68 13 275 68 11 950
Sensor data sheet item number 90 33 511 90 33 510 90 23 924