20
Technical dataen
Cl
2
OV
1)
1) Only for ethylene oxide.
OV-A
1)
Ozone Phosgene
Measuring range 0 ... 20 ppm 0...200ppm 0 ... 200 ppm 0 ... 10 ppm 0 ... 10 ppm
Calibration gas concentration 5ppm inN
2
20 ppm in N
2
20 ppm in N
2
0.5 ... 9 ppm O
3
3.8 ... 9 ppm
Temperature range, operation -30 ... 50 °C
-22...12F
-20 ... 50 °C
-4 ... 122 °F
-20 ... 50 °C
-4 ... 122 °F
-20 ... 50 °C
-4 ... 122 °F
-20 ... 35 °C
-4 ... 99 °F
Alarm threshold A1
2)
2) Observe special, customer-requested settings. Device parameters can be changed with the Dräger CC-Vision Basic PC software.
0.5 ppm 10 ppm 10 ppm 0.1 ppm 0.1 ppm
Acknowledgeable
Latching - - - - -
Alarm threshold A2
1)
1ppm 20 ppm 20 ppm 0.2 ppm 0.2 ppm
Acknowledgeable - - - - -
Latching
TWA threshold value
1)
0.5 ppm - - 0.1 ppm 0.1 ppm
STEL threshold value
1)
0.5 ppm - - 0.1 ppm 0.1 ppm
Mean value period 15 min - - 15 min 15 min
Measurement accuracy
Zero point: ±0.05 ppm ±3 ppm ±5 ppm ±0.02 ppm ±0.01 ppm
Sensitivity: [% of the measured value] ±2 ±5 ±20 ±3 ±10
Long-term drift (20 °C)
Zero point: ±0.2 ppm/a ±5 ppm/a ±5 ppm/a ±0.02 ppm/a ±0.2 ppm/a
Sensitivity:
[% of measured value/month
] ±2 ±2 ±3 ±2 ±2
Sensor item number
3)
3) The sensors have a limited life span. Excessive storage periods compromise the operational life time of the sensors. The adequate temperature range for storage is 0 ... 35 °C (32 ... 95 °F).
68 10 890 68 11 530 68 11 535 68 11 540 68 12 005
Sensor data sheet item number 90 23 924 90 23 994 90 23 995 90 33 259 90 23 924