40
Maintenance
en
6.1 Maintenance intervals
Only use authentic Dräger parts.
Check for correct function and tightness again after
maintenance work and/or replacement of components.
Unit part Work to be performed
Before
each use
After
each use
Every year
Every
2 years
Every
10 years
Diving regulator
(complete)
Check functions and air-tightness X X X
1)
Cleaning X
Complete overhaul by manufacturer X
1)
Pressure reducer Visually inspect O-rings for damage
(cylinder coupling)
X
Replace O-rings X
1)
Lung demand valve Disinfect
2)
, flush X
3)
X
Visually inspect diaphragm,
exhalation valve and O-rings
X
Replace diaphragm, exhalation valve and
O-rings
X
1)
Medium-pressure
hose
Test X
1)
Replace X
1)
1) Only by Dräger personnel or qualified specialists.
2) Use e.g. Incidin
®
Rapid.
3) In case of user change.