Replacement parts
The two rubber O-rings in the hose connector should be replaced annually to reduce malfunction.
Remove the old O-ring using the removal tool and insert a new O-ring in the groove (Fig 9).
Check the ow rate
With a lter and hose attached, turn on and wait for the self-test to complete. Cover the open end
of the breathing hose with your hand. After 5 seconds you should hear the unit start to run faster and
an alarm will be triggered. The panel will ash red indicating the air supply is blocked (Fig 8). If no
alarm is triggered, the fan speed remains unchanged, or the warning does not stop after removing
your hand - do not use and consult your supplier.
Dräger X-Plore
®
8000 Cleaning
and Maintenance Guide (continued)
Fig. 8 The ow rate indicator
Fig. 9 Location of the O-ring
THE RPE EXPERTS

Other manuals for Dräger X-plore 8000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger X-plore 8000 and is the answer not in the manual?

Dräger X-plore 8000 Specifications

General IconGeneral
Product TypePowered Air-Purifying Respirator (PAPR)
Protection ClassTH3
Charging TimeApproximately 4 hours
Air FlowMinimum 170 l/min
DisplayLCD with status information
Filter TypeParticle filters, gas filters, or combination filters (depending on the application)
AlarmVisual and audible alarms for low battery
ApprovalsEN 12941
Battery LifeUp to 4-20 hours (depending on battery and operating conditions)

Related product manuals