Guidance and manufacture’s declaration-electromagnetic immunity for all EQUIPMENT and SYSTEMS
Guidance and manufacture’s declaration-electromagnetic immunity
The MQ3000 Pulse Oximeter is intended for use in the electromagnetic environment specified specified below.
The the user of MQ3000 Pulse Oximeter should assure that it is used in such an environment.
Immunity test IEC60601 test level Compliance level Electromagnetic environment-guidance
Electrostatic dis-
charge (ESD) IEC
61000-4-2
±6KV contact
±8KV air
±6KV contact
±8KV air
Floors should be wood, concrete or ceramic tile.
If oor are covered with synthetic material, the
relative humidity should be at least 30%.
Power frequency
(5 0Hz) magnetic
eld IEC 61000-4-8
3A/m 3A/m Power frequency magnetic elds should be at
levels characteristic of a typical location in a
typical commercial or hospital environment