8 | Maintenance
172
D-R 290
Se
qu.
no.
Simple fault Cause;
(●) available actions by whom
[140] Stepper motor error The stepper motor driver has reported a fault.
Check the stepper motors for correct operation,
check that plug X4 is securely seated
If the error occurs repeatedly or over an exten
ded period, the electronics may be defective
Specialised
personnel
[141] Stepper motor error The stepper motor driver has reported a fault.
➔ Check the stepper motors for correct operation,
check that plug X4 is securely seated
If the error occurs repeatedly or over an exten
ded period, the electronics may be defective
Specialised
personnel
[142] Purge air flow 1 (measuring
head) too low
Purge air flow is below the fault threshold. Possible
causes: Blower defective, filter clogged, purge air
hose split, parameterisation of the purge air inlet
defective.
It can be expected that the measurement func
tion may be impaired by dust in the welded-in
pipes, by a rapid increase in contamination or
damage to the device due to high temperatures.
Check the purge air flow / restore the flow or re
move the device from the flue stack.
Specialised
personnel
[143] Purge air flow 2 (reflector) too
low
Purge air flow is below the fault threshold. Possible
causes: Blower defective, filter clogged, purge air
hose split, parameterisation of the purge air inlet
defective.
It can be expected that the measurement func
tion may be impaired by dust in the welded-in
pipes, by a rapid increase in contamination or
damage to the device due to high temperatures.
Check the purge air flow / restore the flow or re
move the device from the flue stack.
Specialised
personnel
[144] Fail-safe shutter 1 (measuring
head) closed
The fail-safe shutter at the measuring head is
closed. This means that valid measurement values
are no longer being reported.
Check the reason for closure of the shutter (has
the purge air failed?)
Operating
personnel
[145] Fail-safe shutter 2 (reflector)
closed
The fail-safe shutter at the reflector is closed. This
means that valid measurement values are no lon
ger being reported.
Check the reason for closure of the shutter (has
the purge air failed?)
Operating
personnel
Table 8.4: List of simple faults for the D‑R 290
Se
qu.
no.
Critical fault Cause;
(●) available actions by whom
[192] Hardware fault The electronic hardware of the device has a serious
internal fault.
Exchange the electronics (No. 22)
Service technician
[193] Hardware fault The electronic hardware of the device has a serious
internal fault.
Exchange the electronics (No. 22)
Service technician