Handheld Pulse Oximeter (Reorder No. 36403) 31
7. ALARM SETTING (Continued)
8. BEEP ON/OFF
E 11: The value of pulse rate is lower than the low limit.
E 12: The value of pulse rate is higher than the high limit.
E 13: Power supply is insufcient.
When E9, E10, E11 or E12 occurs, the value of
parameterwillash.
• Check the error code.
• Press the right key under normal screen, it will indicate an
error code. To return to normal screen please press menu
button again.
WARNING!
When alarm occurs, DO as follows:
• Check which type of alarm is on.
• Check patient’s condition if parameter alarm is on.
• Make the alarm mute if necessary.
• Check whether the alarm is in proper condition in case
of no warning.
Press the menu key ve times under the normal screen to enter
access to this function, the associated screen will be displayed.
Press up or down key to set the pulse beep on/off.