Handheld Pulse Oximeter (Reorder No. 36403) 35
9. MAINTAIN & CLEANING (Continued)
9.2 CalibrationandVerication
The performance should be checked every year and after
any kind of maintenance or repair.
Required Test Equipment: SpO
2
signal simulator.
NOTICE: The simulator cannot be used to assess the
accuracy of a Pulse Oximeter probe or a Pulse Oximeter.
9.2.1ControlKeyVerication
Press Menu key to display the history data.
9.2.2SoundVerication
a. Set the Pulse Oximeter sound ON.
b. You hear the simulated heart beep sound.
9.2.3 SpO
2
&PulseRateMeasurementValueVerication
a. Connect SpO
2
Probe to the SpO
2
connector on the
oximeter.
b. Insert the operator’s nger into the nger sensor, the
SpO
2
measured value of a healthy person should be
from 95% to 99%, and the pulse rate is the same as a
heart rate.
c. If SpO
2
Simulator is available, verify the accuracy of
Oxygen Saturation Value with probes as follows:
Oxygen Saturation Tolerance
96% ±2%
86% ±2%
70% ±3%