42 Handheld Pulse Oximeter (Reorder No. 36403)
APPENDIX A SPECIFICATIONS (Continued)
APPENDIX B
Symbols
minus; negative © copyright
% percent / per; divide; or
+ plus = equal to
< less than > greater than
less than or equal to greater than or equal to
± plus or minus × multiply
Guidance and manufacturers declaration –
Electromagnetic Immunity
The Handheld Pulse Oximeter is intended for use in an electromagnetic
environment specied below. The customer or the user of the Pulse
Oximeter should assure that it is used in such an environment.
Immunity Test IEC 60601
Test level
Compliance
Level
Electromagnetic
environment guidance
Electrostatic
Discharge
(ESD)
IEC61004-2
6kV contact
8kV air
6kV contact
8kV air
Floors should be wood,
concrete or ceramic tile.
If oor are converted
with Synthetic material,
the relative humidity
should be at least 30%