TRANSDUCER INSTALLATION
The transducers for the TFXL meter contain piezoelectric crystals that transmit and receive ultrasonic signals through the
walls of liquid piping systems.
DTTR, DTTN and DTTH transducers are relatively simple and straightforward to install, but spacing and alignment of the
transducers is critical to the system’s accuracy and performance. CAREFULLY EXECUTE THESE INSTRUCTIONS.
DTTS and DTTC small pipe transducers have integrated transmitter and receiver elements that eliminate the requirement for
spacing measurement and alignment.
Mounting the DTTR, DTTN and DTTH clamp-on ultrasonic transit time transducers takes four steps:
1. Select the optimum location on a piping system.
2. Select a mounting conguration.
3. Enter the pipe and liquid parameters into the UltraLink software utility or key them into the transmitter. The UltraLink
software utility or the transmitter’s rmware calculates proper transducer spacing based on these entries.
4. Prepare the pipe and mount the transducers.
Mounting Location
Select a Mounting Location
The first step in the installation process is the selection of an optimum location for the flow measurement to be made. For this
to be done effectively, a basic knowledge of the piping system and its plumbing are required.
An optimum location is defined as:
• A piping system that is completely full of liquid when measurements are being taken. If the pipe may become completely
empty during a process cycle an error code 0010 (Low Signal Strength) will be displayed on the transmitter while the
pipe is empty. This error code will clear automatically once the pipe refills with liquid. Do not mount the transducers in an
area where the pipe may become partially filled, such as the highest point in a flow loop. Partially filled pipes will cause
erroneous and unpredictable operation of the transmitter.
• A piping system that contains lengths of straight pipe such as those described in Table 1 on page 16. The optimum
straight pipe diameter recommendations apply to pipes in both horizontal and vertical orientation. The straight runs in
Table 1 apply to liquid velocities that are nominally 7 fps (2.2 mps). As liquid velocity increases above this nominal rate, the
requirement for straight pipe increases proportionally.
• An area where the transducers will not be inadvertently bumped or disturbed during normal operation.
• NOT on downward flowing pipes unless adequate downstream head pressure is present to overcome partial filling of or
cavitation in the pipe.