UltraLink Error Codes
Revised 04-06-2015
Warnings
Code Description Correction
0001 Serial number not present
Hardware serial number has become inoperative. System
performance will not be influenced.
0010
Signal Strength is below Signal Strength Cutoff
entry
Low signal strength is typically caused by one of the
following:
» Empty pipe
» Improper programming/incorrect values
» Improper transducer spacing
» Non-homogeneous pipe wall
Removing the resistors from the transducer terminal block
can boost the signal.
0011
Measured speed of sound in the liquid is greater
than ±10% of the value entered during
transmitter setup
Verify that the correct liquid was selected in the BASIC menu.
Verify that pipe size parameters are correct.
Class C
Errors
1001 System tables have changed
Initiate a transmitter RESET by cycling power or by selecting
SYSTEM RESET in the SEC MENU.
1002 System configuration has changed
Initiate a transmitter RESET by cycling power or by selecting
SYSTEM RESET in the SEC MENU.
Class B
Errors
3001 Invalid hardware configuration Upload corrected file.
3002 Invalid system configuration Upload corrected file.
3003 Invalid strategy file Upload corrected file.
3004 Invalid calibration data Re-calibrate the system.
3005 Invalid speed-of-sound calibration data Upload new data.
3006 Bad system tables Upload new table data.
Class A
Errors
4001 Flash memory full Return transmitter to factory for evaluation
Table 6: TFXL error codes