1. REFER TO TRANSMITTER'S INSTALLATION MANUAL FOR TRANSDUCER LOCATION AND MOUNTING INSTRUCTIONS.
2. WARNING - TO PREVENT IGNITION OF FLAMMABLE ATMOSPHERES, DISCONNECT POWER BEFORE SERVICING.
3. WARNING: SUBSTITUTION OF COMPONENTS MAY IMPAIR INTRINSIC SAFETY.
4. NO REVISION TO DRAWING WITHOUT PRIOR CSA-INTERNATIONAL APPROVAL.
5. ASSOCIATED APPARATUS MANUFACTURER'S INSTALLATION DRAWING MUST BE FOLLOWED WHEN INSTALLING THIS EQUIPMENT.
6. INSTALLATION IN CANADA SHOULD BE IN ACCORDANCE WITH THE CANADIAN ELECTRICAL CODE, CSA C22.1, PART 1, APPENDIX F.
NATIONAL ELECTRICAL CODE (ANSI / NFPA 70) SECTIONS 504 AND 505 AND THE
ANSI / ISA RP12.6 INSTALLATION OF INTRINSICALLY SAFE
8. THE MAXIMUM NON-HAZARDOUS LOCATION VOLTAGE IS 250V AC/DC.
7. INSTALLATION SHALL BE IN ACCORDANCE WITH THE
SYSTEMS FOR HAZARDOUS (CLASSIFIED) LOCATIONS.
FOR INCIDENTAL AND
TEMPORARY SUBMERSION
D002-1401-003
CONDUIT SUITABLE
FLEXIBLE ARMORED
990' MAX.
(302 METERS)
(RG-59/U COAX, BELDEN 9463, OR
BELDEN 9463DB OR EQUAL ONLY)
SENSING SURFACE: COUPLE TO
PIPE WITH RTV OR SILICONE
GREASE SUPPLIED, PER INSTALLATION
MANUAL TFXD O&M.
3' MIN.
(0.93 METERS)
(2 PLACES)
DYNASONICS DTT SERIES
TRANSDUCERS
MODEL NO: DTTN-xxx-Axxx-F
CLASS I, DIVISION 1 GROUPS C AND D
MAXIMUM AMBIENT TEMPERATURE: -40° TO +85°C
HAZARDOUS (CLASSIFIED) LOCATION
MAXIMUM AMBIENT TEMPERATURE: -40°C TO 50°C
TRANSMITTER PER
CONNECT TO
INSTALLATION
MANUAL
NON-HAZARDOUS LOCATION
TEE FITTING
D002-1201-002
SEAL OFF CONDUIT
PER INSTALLATION
NOTES 6 & 7
B
1
2
B
A
C
D
E
F
G
H
A
B
C
D
E
F
G
H
1
345610 9 8 712 11
12
11
10 9 8
7
6
5
4 3
2
1
THIS DRAWING IS PROPRIETARY TO RACINE FEDERATED INC. RECEIPT OR POSSESSION CONFERS NO RIGHT TO USE
THE SUBJECT MATTER OF THIS DRAWING OR TECHNICAL INFORMATION SHOWN; NOR THE RIGHT TO REPRODUCE THIS
DRAWING OR ANY PART EXCEPT FOR THOSE SUPPLIERS OF RACINE FEDERATED INC. WHO RECEIVE A WRITTEN
REQUEST FOR MANUFACTURE OR SIMILAR USE.
PART NUMBER:
D
59380
THIS DRAWING WAS DONE ON AUTOCAD AND
CAN ONLY BE REVISED ON AUTOCAD SYSTEM.
ANY MANUAL CHANGES DONE TO THIS DRAWING
WILL BE IGNORED UNLESS AUTHORIZED.
NOTES: UNLESS OTHERWISE SPECIFIED
ZONE REV DESCRIPTION E.C.N. D ATE APPROVAL
REVISIONS
UNLESS OTHERWISE SPECIFIED
DIMENSIONS ARE IN INCHES
TOLERANCE ON DECIMALS
SUPERSEDES:
DATE:
CHECKED BY:
DRAWN BY:
ENGINEER:
MATERIAL:
SCALE: SHEET:
CODE I.D. NO.SIZE: PART NUMBER:
CURRENT REV:
3. RADII TO BE .005/.010
2. FINISH TO BE 63
1. REMOVE ALL BURRS AND BREAK SHARP EDGES .005/.010
.00 ±.010, .000 ±.005
ANGLES ±1/2°
REVISED BY:
NAME:
TEL: 262-639-6770 FAX: 262-639-9857
RACINE, WISCONSIN U.S.A.
DIVISION OF RACINE FEDERATED, INC.
SEE ABOVE
T. PAUL
11/28/01
N.GARMON 12/23/11
CONTROL DRAWING
I.S. BARRIER & DTT TRANSDUCER
D091-1053-005
D091-1053-005
NONE C 2 OF 2
ALL A
UPDATED PER CSA
- 11/11/04
C6 B
1.) ADDED BELDEN 9463 & 9463DB
R1005 10/25/10
OR EQUAL
E3 C 1.) PICTORIAL CHANGES TO SHOW R1412 12/23/11
CORRECT ARTWORK ON DECAL
DYNASONICS I.S. BARRIER
MODEL D070-1010-002
RED/BLUE
BLACK/CLEAR
RED/BLUE
BLACK/CLEAR
I.S. MODULE
PART NO. D070-1010-001
I.S. WIRING
10' MAX
3.60
(MTG. HOLES)
6.26
(MTG. HOLES)
C
1
SAFE HAZ
HAZ
SAFE
MODEL: 070-1010-002 I.S. Barrier-Ultrasonics
D003-0905-117
Figure 44: Control drawing, IS barrier DTT transducers flexible conduit
Control Drawings
Page 53 April 2016 TTM-UM-00369-EN-03