providing a window into the process
will need to be installed before safe operation of the machine is possible. Other issues
important to obtaining accurate data must also be addressed before valid testing can begin.
Level the Melt Indexer
Using a small round bubble level, level the melt indexer. Place the level on top of the COLD
barrel and using the adjustable screw feet, bring the machine into level. Tighten the locking nuts
to keep the feet in level position Test shake the melt indexer for stability.
Some companies may choose to bolt the machines directly to their benches.
This can be done by removing the levelling feet and using their mounting holes to mount the
instrument to a bench or table. Use caution and remove power before doing so. Mounting this
way will entail removing some covers for internal access to the internal, main instrument plate.
Be sure to remove the level before turning on the machine
. The level will be damaged if it gets
hot.
P/N: M0726(3) | Rev: 012020 | ECO: 51182
www.dynisco.com