EasyManua.ls Logo

Eagle EI-2000 - 4-20 mA Output Setup and Testing

Eagle EI-2000
44 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
EI-1000 & 2000: Setup, Calibration and Troubleshooting.
Page 12 of 19
4-20 mA parameter setup:
Enter the configuration as shown on page 9.
Use the UP arrow to get to P600. This sets what the 4-20 output follows.
The selections are: Gr - gross weight, rE - remaining weight, GT - gross
total, rT - remaining total. “Total” modes transmit the sum of both sides of
the display on channel 1.
EI-1000 - Press SELECT and use the arrows to change the value if
needed. Press SELECT again.
EI-2000 - Use the right arrows to change the value if needed.
Now go to parameter P610. This set the weight value that equals 20 mA.
Change the value if needed as shown above.
See “4-20 mA output testing” for calibrating or testing if needed.
4-20 mA output testing:
EI-1000: Press and hold [SELECT] + [RIGHT ARROW] for greater than three seconds.
EI-2000: Press and hold [LEFT SELECT] + [RIGHT SELECT] for greater than three seconds.
Use the arrows to get to “AOUT”. Press SELECT.
“Ao.test” (EI-1000) or “Ao T 4.00” (EI-2000) will be displayed.
If using an EI-1000 press SELECT and “4.00” will start to flash.
The output is now forced to 4 mA. Connect a current meter or check the
remote readout to check the output.
Using the arrows (right side on EI-2000) will force the output to the value
shown.
Go up to 20.0. Check the output is 20mA or full scale on the remote
readout.
Go to 12.00 mA. Check the output is 12mA or half scale on the remote
readout.
On the EI-1000 press SELECT when done to return to Ao.TEST.
Calibrating the 4-20mA outputs:
Enter the configuration as shown above and select “AOUT” as shown in the first step.
1. Go up to “Ao.1 4“ (“Ao.1 4.0” on EI-2000, use the left arrows.)
2. Connect an mA/current meter capable of displaying 20.00mA to
0.01 mA to CH1 output.
3. On the EI-1000 press SELECT. The “4” will flash.
4. Use the up/down arrows to set the output to 4.00 mA on your
meter. The readout on the indicator will not change. (Use right
arrows on EI-2000)
5. On the EI-1000 press SELECT. The “4” will stop flashing.
6. Use the up arrows (left on EI-2000) to get to “A.o 1 20”
7. Press select on the EI-1000.

Other manuals for Eagle EI-2000

Related product manuals