User manual SEFELEC 5x Series 105 v1.06
NOTE
Press the Start push button. For safety reasons, it is
necessary to press the
button for > 1 second for the device to switch into measurement mode. If the
button is pressed for too short a time a message will be displayed momentarily
on the LCD screen.
The measurement cycle comprising as a maximum 7 phases, takes place according to Rise,
Hold and Fall times with the display of the voltage and current according to the following screen:
At the end of the test the display shows if the measurement is between Imin and Imax i.e. con-
sidered PASS.