  PRECAUTIONS TO BE OBSERVED DURING MEASUREMENT: 
 
WARNING
 
Never disconnect a capacitive sample without switching to the dis-
charge mode and waiting for the time required for the capacity of the 
sample  to  be  discharged  into  the  resistance  of  1.5  MΩ  of  the  dis-
charge circuit. 
 
 It is recommended to check the proper functioning of the device regularly by 
connecting an element of known value to the measurement output and perform-
ing a measurement. In the absence of an element of known value, it is possible 
to make a measurement on an open circuit and on a short-circuit to confirm the 
consistency of the indications of the device. 
After having connected the component for measurement applying the precautions described 
above and selected the measurement settings required, return to the measurement screen: