User manual SEFELEC 5x Series  183   v1.06 
 
FILT (FILTer) NOR/CAP  
Enables the normal or the capacitor measurement modes. 
For instance : CONF:FILT CAP 
 
DET (DETection) OFF / I / I+DELTA / DELTA :  
Hipot function only sets the mode of the breakdown. 
 
IDE (IDElta) <NR3> 
Hipot function : Delta I breakdown detection value adjustment 
 
MKM (MEGOHMKM) ON OFF 
Allows to activate or disactivate the Megohm.km mode  
 
 
MKMLEN (MEGOHMKMLENGTH) <NR1> 
with <NR1> cable length (1 to 16000 meters) in Megohm.km mode  
 
MKMCOUNT  (MEGOHMKMCOUNT) <NR1> 
With <NR1> number of wires (0 to 999) in Megohm.km mode  
 
MEAS (MEASure)   
Runs the current function. 
 
STOP   
Stops the current function. 
 
QUIT  
Exits the current function. 
 
SRQ   
Similar to the Service Request feature of the IEEE488-2 bus, this code allows the 
emission of the Z character to inform the computer about the events (end of test, 
format error, interlock open, …). This command has to be sent at the program 
start, after the REM command. 
 
18.1.5.  REQUEST DEPENDING ON THE DEVICE 
 
MEAS?  
Return the current measurement value(s) regarding the running function. 
 
Insulation     : VOLT 9.900E+01 AMP 2.000E-06 OHM 4.950E+06 
Dielectric strength  : VOLT 1.000E+03 AMP 7.000E-05 AMP 0.100E-06 
Continuity    : VOLT 2.810E+00 AMP 1.000E+01 OHM 2.810E-01