FCA22InstructionManual
Page6
Figure2.2
Temperature
TemperaturevariationinfluencestheFCSby changingthepermeabilityofthePTFEmembraneandthe
Nernstianresponseofthesensor.Combinedthesechangesaccountforachangeof4%/C°.Thechange
follows the temperature, as the temperature increases the output of the sensor increases, as the
temperature drops the outpu
t drops. The FCA22 automatically compensates for the changes. The
temperature sensor is located in the pH sensor and it has a response time of several minutes. Rapid
changes of temperature will introduce an error until the sensor has equilibrated to the new
temperature.Calibrationshouldbedoneclosetotheprocesstemperatureforthehighestac
curacy.Se e
fig.2.3.
Figure2.3
0
20
40
60
80
100
0 102030
%output
Flowrate,gal/hr
FCAFlowdependence
0
10
20
30
40
50
0204060
mV/ppm
Cl2
Temperature
FCATemperature
Dependence,4%/C°