Troubleshooting
The FCA22 was evaluated and calibrated at the factory before shipment. Upon initial start up the
systemshouldrequireminimaltonoadjustments.
Verify the system has ade quate flow, greater than 10 gals /hr, the pH electrode and the temperature
sensor arereadingcorrectly. These parameters effect the measurementand must meet the standards
listed in th
e Calibration Section above. If these conditions are met and problems still exist use the
TroubleshootingTabletofindaremedy.
TroubleshootingGuide
Symptom PossibleCause Remedy
DisplayedvalueisHigher
thanDPDtestvalue.
InsufficientPolarizationtime
DamagedMembraneCap
pHindicationhigherthan
actualvalue
Temperatureindication lower
thanactualvalue
Electricalshortorwet
connectioninsidethesens or
orcableassembly
PolarizeFCSforfull90
minutesbeforecalibration.
ReplaceMembraneCap
CalibratepHsensororadjust
valueonun
compensatedFCA
CalibratetheTemperature
(seeCalibration)The
temperaturesensorlagsthe
processtemperaturewaitfor
temperatureequilibrium.
Removemeasuringchamber
anddrythecathodesurface,if
theindicationdoes notgoto
zerothereisleakage.Replace
thesensor.
DisplayedvalueisLowerthan
DPDtestvalue
Chloramineorotheroxidants
presentinsampleyieldinga
highDPDtest.
Coatedordirtymembrane
Replacewaterandchlorinate
Cleanorreplacethe
membrane