What to do if ecodyst Laboratory Equipment has a power failure?
- NNicole RamirezAug 10, 2025
If your ecodyst Laboratory Equipment experiences a power failure, first, check that the power cord is properly connected to the controller and that the electricity supply is both suitable and regulated. Also, check if the mantle temperature has risen too high, causing a temperature overshoot. If so, allow the temperature to decrease below the high limit before pressing the reset button on the temperature controller to restart. Finally, make sure the setpoint temperature is not set higher than the high limit temperature.