SE-1515 PC ECG User Manual Preparations Before Operation
- 31 -
N
RL
Right leg/Upper leg as close to torso as
possible
NEHB Placement
IEC
AHA
Electrode Placement
N
st
/C3R
A1/V3R
Attachment point of the second rib to the right sternal edge
N
ax
/C4R
A2/V4R
Fifth intercostal space on the left posterior axillary line
N
ap
/C4
V4
Left mid-clavicular line in the fifth intercostal space
V3R+V4R +V5R (Right)
IEC
AHA
Electrode Placement
C3R
V3R
Right anterior chest opposite of C3
C4R
V4R
Right anterior chest opposite of C4
C5R
V5R
Right anterior chest opposite of C5