SE-1515 PC ECG User Manual ECG Analysis
- 68 -
5.4 HRV
The HRV ECG analysis screen includes two tabs: Auto Diagnosis result and Waveform.
NOTE:
1. The HRV sampling time can be set in the Sample Setting window.
2. The HRV analysis lead can be selected in the Sample Setting window.
Total beat number during the measuring course
Ratio of Maximum RR interval to Minimum RR interval
Standard Deviation of Normal to Normal Intervals
Root Mean Square Successive Difference
NN50
(the total beat number)
The number of duration difference that is more than 50ms between
the adjacent NN durations.
NN50 divide the total NN number
Standard deviation of successive differences between adjacent
normal cycles
The triangular interpolation of NN interval histogram
Total NN number divide the NN intervals of the highest percentage
Ratio of low frequency to high frequency