SE-1515 PC ECG User Manual ECG Sampling
- 42 -
4.4.1.4 Sample Setup
Click on Setting on the resting ECG sampling screen, and the System Setting window will be
displayed. In the System Setting window, click Sampling Setting to enter the Sampling Setting
window.
Options provided are DP12, DX12, DE15, SE12, DE18, and DEMO.
If DX12 is selected, you can view the address of the DX12 wireless receiver
after clicking on Receiver Address.
You can configure the COM ports used for transmission.
You can configure the Device ID. The device number must be within 30
characters.
WARNING
The device number cannot be changed. Contact the manufacturer if you want to make
any modifications on it.
You can configure the sampling time for resting ECG.
Options provided are 9-lead, 12-lead, 15-lead, 16-lead, and 18-lead
The 9-lead mode is normally used for pediatric ECG or physical examination.
Under the 9-lead mode, you can choose Physical mode.
І, II, ІІІ, aVR, aVL, aVF, V1, V3, V5
Under the 12-lead mode, you can choose from Standard and Cabrera.
І, II, ІІІ, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
aVL, І, -aVR, II, aVF, ІІІ, V1, V2, V3, V4, V5, V6
Under the 15-lead mode, you can choose from Standard+Right, Standard+Back,
Standard+NEHB, Standard+XYZ, and Children mode.
І, II, ІІІ, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6, V3R,
V4R, V5R
І, II, ІІІ, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6, V7,
V8, V9