SE-601 Series Electrocardiograph Service Manual Inspection and Troubleshooting
- 40 -
TP14
+2.0V±0.10V
TP1
+7.75V±0.75V
TP3
+28V±1.40V, see NOTE1
GND
GND
TP19
6.95~-6.85V, see NOTE2
TP12
+1.8V±0.09V
NOTE1: Only test TP3 on SE-601C with the AUO 5.7” V0 display.
NOTE2: Only test TP19 on SE-601A. If display problem occurs and the voltage
exceeds the range, you can adjust potentiometer VR2 to make TP19 meet the
requirements by using the debugging pen.
4) Compare the measurement results with the reference results in the list. If any one of the
results exceeds the reference range, the main board defection is confirmed. Replacement of
the main board is recommended.
TP12 TP19 GND TP3 TP1
TP14 TP5 TP6 TP8 TP7
理邦保密文件
EDAN CONFIDENTIAL