SE-601 Series Electrocardiograph Service Manual Inspection and Troubleshooting
- 45 -
5.3.6 Display Drive Board Test
NOTE: Only test the display drive board on SE-601B/C.
To verify the display drive board,
1) Open the main unit following the procedures described in Chapter 6, Electrocardiograph
Disassembly.
2) Switch on the device.
3) Measure the voltage to earth of the test points listed below by using a multimeter.
Item
Test Point
Reference Result
TP1
8.40~10.50V
TP3
3.14~3.46V
TP7
4.75~5.25V
TP8
-10.50 ~ -9.50V
TP9
14.30~15.70V
VR1_2
1.80~2.20V
FB2_2
5.00~556V
理邦保密文件
EDAN CONFIDENTIAL