SE-601 Series Electrocardiograph Service Manual Electrocardiograph Disassembly
- 56 -
3) Pull out the fuse socket from the fuse socket slot.
4) Remove the old fuse. Insert the new fuse using the same specification (T3.15AH250V,
Ø20mm) supplied by the manufacturer.
5) Push the fuse all the way into the fuse socket slot.
WARNING
Ruptured fuses must only be replaced with those of the same type and rating as the
original.
6.2 Internal Boards and Interfaces
1. ECG Board -- Receiving and Processing ECG Signals
J4-- connected to main board J15.
Pin
Description
2
GND
3, 4
Data transmitting
5
Vcc +5V
理邦保密文件
EDAN CONFIDENTIAL