© Edwards Limited 2018. All rights reserved. Page 5
Edwards and the Edwards logo are trademarks of Edwards Limited.
Technical Data
D023-92-880 Issue F
2Technical Data
2.1 Mechanical data
2.2 Performance, operating and storage conditions
2.3 Electrical data
Note: The electrical power consumption of the APGX-H is reduced when the system is at vacuum. The relationship
between the output signal voltage and pressure is described in Section 4.1.
Dimensions Refer to Figure 2
Mass
APGX-H NW16 ALI 110 g
APGX-H NW16 ST/ST 160 g
APGX-H NW25 ST/ST 170 g
APGX-H
1
/
8
NPT ST/ST 150 g
Volume of gauge tube 11 cm
3
Enclosure rating IP40
Ambient temperature
Operation +5 to +60°C
Storage -30 to +70°C
Ambient humidity (operation) 80% RH up to 31°C decreasing linearly to
50% at 40°C and above
Operating conditions 2000 m (indoor use only)
Maximum internal pressure 10 bar absolute (9 bar gauge)
Pressure range 1333 to 3 x 10
-4
mbar (1000 to 2.3 x 10
-4
Torr)
Filament operating temperature (above ambient) 100°C
Pollution degree 2
Electrical supply
Type NEC Class 2 or Limited Power or Limited Energy
Voltage +14.5 V to +30 V d.c.
Maximum power consumption 1.5 W
Start-up current 120 mA
Electrical connector FCC68 / RJ45 type, 8-way
Pressure output signal
Range 1.9 V to 9.125 V d.c.
Error range output < 1.8 V d.c. or output > 9.2 V d.c
Min load impedance 10 k
Max output current 1 mA
Set-point
Adjustment range 1.8 V to 9.3 V