Appendix A
58 Incorrect pressure format A command had a wrongly formatted pressure. Check format
specified in Appendix B.
81 No gauge connected An attempt was made to read or control a gauge when it was
not connected. Check that gauge number is correct.
82 Unknown gauge type A gauge is connected that cannot be recognised. Check that
the gauge is a type listed in Table 1. If not, a software
upgrade may be needed. See Section 5.4.
83 Gauge not reading pressure An attempt was made to read a gauge which was switched
off. Check that the gauge is switched on. Use the command
!GW if necessary.
84 AIM gauge striking An attempt was made to read a gauge which was striking.
Wait until the gauge has struck and try again.
90 Incorrect gauge type, query/command
not appropriate
The command or query is not supported by the particular
gauge e.g. an attempt was made to switch an APG on or off.
91 Gauge turn-on is inhibited by link A "GW1=1" command has been sent when 2 ADCs are linked,
and the pressure is too high to allow the controlled gauge to
turn on.
Table A1 - ADC error numbers (continued)
ERR
number
Meaning Possible cause / remedy