4.6.2 Check the direcon of rotaon
CAUTION:
Ensure that the pump-motor rotates in the correct direcon. If it does not, the pump
and the vacuum system can become pressurised.
Refer to Figure: General view of the pump on page 11.
1.
Watch the motor-cooling fan through the fan cover on the end of the motor.
2.
Switch on the electrical supply to the motor for a few seconds.
3.
Check that the motor-cooling fan rotates in the direcon shown by the arrow on
the end of the oil box. If the rotaon direcon is incorrect, immediately switch o
the electrical supply and contact us or your supplier for advice.
4.7 Inlet-port connecon
4.7.1 General requirements
To connect the pump to the vacuum system, use either the inlet-port on the top of the
pump (Figure: General view of the pump on page 11) or the inlet-port on the side of the
pump (not shown). To use the side inlet-port, refer to the procedure in Side inlet-port
connecon on page 21.
When using the top or the side inlet-port:
▪
Connect to the NW10 ange supplied on the pump, use the centring ring and
O‑ring supplied.
▪ Remove the NW10 ange and connect to the 3/8 inch BSP threaded hole.
▪ Remove the NW10 ange and replace it with a NW25 ange adaptor (available as
an oponal accessory, see Spares and accessories on page 36) and then connect to
the NW25 ange.
Take note of the following informaon when connecng the pump to the vacuum
system. Refer to Spares and accessories on page 36 for details of the accessories
menoned below.
▪
For opmum pumping speeds, ensure that the pipeline connected to the inlet-port
is as short as possible and has an internal diameter not less than the inlet-port
diameter.
▪
Support the vacuum pipelines to prevent loading of the coupling‑joints.
▪
If necessary, incorporate exible bellows in the system pipelines to reduce the
transmission of vibraon and to prevent loading of coupling‑joints. If using exible
bellows, ensure that the bellows which have a maximum pressure rang which is
greater than the highest pressure that can be generated in the system. We
recommend you to use manufacturer's exible bellows.
▪
Use a suitable inlet‑lter if pumping condensable vapours or if using the pump for
very dusty applicaons.
▪
Use a suitable valve to isolate the pump from the vacuum system when pumping
condensable vapours or to maintain vacuum when the pump is switched o.
▪
Ensure that sealing surfaces are clean and scratch‑free.