D135-10-880C - Troubleshooting
Page 103
W83
All EEPROM parameter lost.
Check the settings.
EEPROM on back plane is empty and was
initialized with default valves. Enter all parameters
again.
The EEPROM might be faulty when warning comes
up again after power up.
W84
EEPROM parameter initialized.
Check the settings
Software update has been installed
W85
Lost EEPROM parameter!
Check the settings!
Writing access was interrupted. Check all
adjustments.
An software update was done. In this case the
notice can be ignored.
When warning comes up again after powering up
the EEPROM might be faulty.
Table 24 Fault finding (continued)
No. Displayed message Description and possible solutions