D135-10-880C - Description
Page 22
Default settings
The following parameters are set like shown when in the menu of the ELD500 under Settings
> Parameters, Load/Save is chosen.
Scale linear
Display range: 4 decades
Time axis: 32 seconds
LCD inverted off
Background in standby mode: off
Calibration request: Off
Mass: 4 (helium)
Recorder: Leak rate
Volume: 2
Leak rate unit: mbar l/s
Mode: Vacuum
Trigger level 1: 1E-9 mbar l/s
Trigger level 2: 1E-8 mbar l/s
Trigger level 3: 1E-7 mbar l/s
Leak rate external test leak (vacuum): 1E-7 mbar l/s
Leak rate external test leak (sniffer): 1E-5 mbar l/s
Vent delay: 2 seconds
Automatic purge (ELD500 DRY and ELD500
FLEX only)
On
Pressure: mbar
Minimum volume: 0
Beep: On
Maximum evacuation time: 30 minutes
Audio Alarm Type: Trigger alarm
Maximum pressure limit for sniff mode 0.15 mbar
Minimum pressure limit for sniff mode 0.05 mbar
Control location Local
Alarm delay: 30 seconds
Leak rate filter: Auto
Zero: Enable
Vacuum ranges Normal
Upper display limit 1E-5 mbar l/s