D135-10-880C - Operation
Page 73
Time and date
Main menu > Settings > Miscellaneous > Time and Date
Time and date can be changed on two subsequent pages. Refer to Numerical entries for the
description of the entry.
Language
Main menu > Settings > Miscellaneous > Language
The preferred language can be selected. The default setting is english. The following
languages can be chosen: English, German, French, Italian, Spanish, Chinese, Japanese,
Polish, Russian
Calibration request
Main menu > Settings > Miscellaneous > Calibration request
It can be selected whether the operator is reminded of the fact that a calibration may has
become necessary or not. The default value is off.
If the calibration request is switched on, a corresponding message will appear when 30
minutes have elapsed after power on or if the temperature of the ELD500 has changed by
more than 5 °C (9 °F) since the last calibration.
Softkey 5 Service interval fore-pump
Setting service time interval fore-pump
Softkey 7: Service interval exhaust oil filter
Setting service time interval exhaust oil filter
Softkey 8: Service message exhaust oil filter
Setting service message exhaust oil filter
Softkey 3:
Scrolling down to select the language. Press OK to confirm the selected language.
Softkey 7:
Scrolling up to select the language. Press OK to confirm the selected language.
Softkey 3: Off
The calibration request will be switched off.
Softkey 5: ?
Helptext
Softkey 7: ON
The calibration request will be switched on.