4.8.8 Important notes about Auto Tune
The Auto Tune process is not a dynamic chamber pressure control function. It is intended only as a way of removing
variation when the pump is installed.
The Auto Tune process cannot be used to achieve a pump speed higher than full speed. It is not possible to increase
performance using this function.
The Auto Tune settings are pump-specific, they are not stored in the MicroTIM. If the pump is replaced, the Auto
Tune process must be performed on the replacement pump.