31
Black oil
Before autoclaving, the contra angle must be
lubricated.
Attaching the spray nozzle to oil can and contra
angle, press the oil can button more than 3
seconds, till all the black oil flow out from the
head of the contra angle.
7
Packaging
Pack each component in a separate steam-
sterilization pouch.
Check the validity period of pouch given by the
manufacturer to determine the shelf life.
Use pouches which resist to a temperature up
to 141(286°F) and in accordance with EN
ISO 11607.
8
Sterilization
Steam sterilization at 134, at least 6 minutes.
Minimum drying time after sterilization: 10 minutes.
Use only approved autoclave devices according
to EN 13060 or EN 285.
Use a validated sterilization procedure
according to ISO 17665.
Respect the maintenance procedure of the
autoclave device given by the manufacturer.
Use only this recommended sterilization
procedure.
Control the efficiency (packaging integrity, no
humidity, color change of sterilization indicators,
physicochemical integrators, digital records of
cycles parameters).
The sterilization procedure must comply with
ISO 17665.
Wait for cooling before touching.