Immediately after using, wipe gross contaminations
from the components (contra angle, file clip, lip
hook and insulating sleeve), and put them in
container for transportation.
Prepare the components directly after treatment.
● Do not submerge the components or wipe them
with any of the following functional water (acidic
electrolyzed water, strong alkaline solution, or
ozone water), medical agents (glutaral, etc.), or
any other special types of water or commercial
cleaning liquids. Such liquids may result in
metal corrosion and adhesion of the residual
medical agents to the components.
Warning
● Only the components above can be autoclaved.
● Before first use and after each use, sterilize the above components.