● Before autoclaving, the contra angle must be
lubricated.
● Attaching the spray nozzle to oil can and contra
angle, press the oil can button more than 3
seconds, till all the black oil flow out from the
head of the contra angle.
Pack each component in a separate steam-
sterilization pouch.
● Check the validity period of pouch given by the
manufacturer to determine the shelf life.
● Use pouches which resist to a temperature up
to 141℃(286°F) and in accordance with EN
ISO 11607.
Steam sterilization at 134℃, at least 6 minutes.
Minimum drying time after sterilization: 10 minutes.
● Use only approved autoclave devices according
to EN 13060 or EN 285.
● Use a validated sterilization procedure
according to ISO 17665.
● Respect the maintenance procedure of the
autoclave device given by the manufacturer.
● Use only this recommended sterilization
procedure.
● Control the efficiency (packaging integrity, no
humidity, color change of sterilization indicators,
physicochemical integrators, digital records of
cycles parameters).
● The sterilization procedure must comply with
ISO 17665.
● Wait for cooling before touching.