7 Cleaning disinfection and sterilization
7.1 Cleaning disinfection and sterilization of the reusable tip
Reusable
tip
Only the component above can be autoclaved.
Before first use and after each use, clean disinfect and sterilize the above
component.
Sterilization should be no more than 60 cycles
,
After 60 cycles, discard it.
Reprocessing Instructions
Preparat
the Point
of Use:
Before cleaning, disconnect the component from the Helios 500.
Remove gross contaminations from the components with cold water
(<40°C) immediately after use. Don’t use a fixating detergent or hot
water (>40°C) as this can cause the fixation of residuals which may
influence the result of the reprocessing process.
Store the components in a humid surrounding.
Do not submerge the components or wipe them with any of the
following functional water (acidic electrolyzed water, strong alkaline
solution, or ozone water), medical agents (glutaral, etc.), or any other
special types of water or commercial cleaning liquids. Such liquids may
result in metal corrosion and adhesion of the residual medical agents
Safe storage and transportation to the reprocessing area to avoid any
damage and contamination to the environment.
Deconta
mination
The devices must be reprocessed in a disassembled state.
Observe suitable personal protective measures.
Cleaning
Do a manual pre-cleaning, until the components are visually clean.
Submerge the components in a cleaning solution and flush the lumens
with a water jet pistol with cold tap water for at least 10 seconds. Clean
Page 36 / 51