9 Electromagnetic Compatibility Precautions
● This equipment conforms to IEC 60601-1-2: 2014 + AMD1: 2020 EMC
requirements and tests, Medical Electrical Equipment including
CSIPR11:2009+A1:2010 Group 1, Class B
● The equipment should be installed and used according to the EMC information
provided in the attachment
● Guidance and manufacturer’s declaration refer to the attachment
● Portable and mobile RF communications equipment (including peripherals such
as antenna cables and external antennas) should be used no closer than 30 cm
(12 inches) to any part of the Helios 500, including cables specified by the
manufacturer. Otherwise, degradation of the performance of this equipment could
result.
● Use of this equipment adjacent to or stacked with other equipment should be
avoided because it could result in improper operation. If such use is necessary,
this equipment and the other equipment should be observed to verify normal
operation.
● The use of cables or accessories other than those specified, with the exception of
those sold by the manufacturer of the equipment, as replacement parts for
internal components may result in increased emissions or decreased immunity of
the medical equipment.
Cable information:
Cable Name
Cable Length (m)
Shielded or not
Remark
Power cable
1.8
YES
/