Sterility cannot be guaranteed if packaging is open, damaged or
wet.
Check the packaging before using it (packaging integrity, no
humidity and validity period).
Before sterilization, please remove the reusable tip.
The instructions provided above have been validated by the manufacturer of
the medical device as being capable of preparing a medical device for use. It
remains the responsibility of the processor to ensure that the processing, as
actually performed using equipment, materials and personnel in the processing
facility, achieves the desired result. This requires verification and/or validation
and routine monitoring of the process. Likewise, any deviation by the processor
from the instructions provided should be properly evaluated for effectiveness
and potential adverse consequences.
7.2 Cleaning disinfection of the scanner and holder of
scanner
Holder of scanner
before
Before cleaning and disinfecting, make sure the power is off.
Wipe all the exterior surfaces of the components thoroughly with
a cloth lightly moistened with Ethanol (Ethanol 70 to 80 vol%) at
least 3 min, repeat for 5 times.
Wipe all the exterior surfaces of the components thoroughly with
a cloth lightly moistened with Ethanol (Ethanol 70 to 80 vol%) at
least 3 min, repeat for 5 times.