DO NOT remove the cover of any scanner components. The scanner contains no user-
serviceable parts. For any repairs, contact a qualified Eighteeth service technician.
DO NOT replace the cables provided with the scanner with other cables. Doing so may
damage the scanner and adversely affect the safety protection and EMC performance of the
scanner.
Any other equipment not complying with IEC 60601 shall be kept at least 1.5 meters away
from the patient.
If the equipment is faulty, turn it OFF, display an “Out of Service” notice, and contact a
qualified Eighteeth service technician.
Using components, cables and spare parts other than those specified or provided by the
manufacturer of this equipment may impair the safety protection of the scanner and may
result in increased electromagnetic emissions or decreased electromagnetic immunity of
this equipment and result in improper operation.
No modification of this equipment is allowed.
Additional multiple outlet strips or extension cords should not be connected to the system.
When the device breaks down, stop using it immediately and pull out the USB plug from the
workstation. When the device breaks down, the temperature of the applied part may reach
to 43°C, but no higher than 48 °C.
This equipment contains certain materials and chemical compounds incidental to the
manufacture of electrical and electronic equipment, and improper “end-of-life” disposal of
such equipment can result in environmental contamination. Therefore, this equipment
should not be disposed of as ordinary household waste but should instead be delivered to a
designated electrical and electronic waste disposal or recycling center. For further
information on disposing of electrical and electronic waste, contact the cognizant authority
within the jurisdiction.
Do not use the device continuously for more than 10 minutes.
Workstation
DO NOT place the workstation and the peripheral equipment connected to it in the
immediate vicinity of the patient. Leave at least 1.5 meters distance between the patient and
the equipment.
The scanner is only intended to be connected to a workstation that is at least IEC 60950-1 /
IEC 62368-1, or equivalent standards certified. Connecting the scanner to other equipment
may be hazardous.
Leave a sufficient amount of clear space around the workstation to ensure that it is properly
ventilated.
Position the screen to avoid light reflections from internal or external lighting for maximum
image quality and visual comfort.