11 EMC Tables
Page 60 / 66
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment - guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
+/- 8 kV
contact
+/- 2 kV, +/- 4
kV, +/- 8 kV,
+/- 15 kV air
+/- 8 kV
contact
+/- 2 kV, +/- 4
kV, +/- 8 kV,
+/- 15 kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30 %.
Electrical fast
transients/bursts
IEC 61000-4-4
±2kV
100kHz
repetition
frequency
±2kV
100kHz
repetition
frequency
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC 61000-4-5
Line to line:
±0.5kV, ±1kV
Line to earth:
±0.5kV, ±1kV,
±2kV
Line to line:
±0.5kV, ±1kV
Line to earth:
±0.5kV,
±1kV, ±2kV
Mains power quality
should be that of a
typical commercial or
hospital environment.

Related product manuals