11 EMC Tables
Page 61 / 66
Voltage dips
IEC 61000-4-11
Voltage interruptions
IEC 61000-4-11
0% UT; 0.5
cycle
at 0°, 45°, 90°,
135°, 180°,
225°, 270°,
and 315°
0% UT; 1
cycle and 70%
UT; 25/30
cycles
sine phase at
0% UT;
250/300 cycle
0% UT; 0.5
cycle
at 0°, 45°, 90°,
135°, 180°,
225°, 270°,
and 315°
0% UT; 1
cycle and 70%
UT; 25/30
cycles
sine phase at
0% UT;
250/300 cycle
Mains power quality
should be that of a
typical commercial or
hospital environment. If
the user of devices
require continued
operation during power
mains interruptions, it is
recommended that
devices be powered form
an uninterruptible power
supply or a battery
Rated Power frequency
magnetic field IEC
61000-4-8
30 A/m
50Hz or 60Hz
30 A/m
50Hz or 60Hz
Power frequency
magnetic field should be
at levels characteristic of
a typical location in
a typical commercial or
hospital environment.
Note: UT: rated voltage(s); E.g. 25/30 cycles means 25 cycles at 50Hz or 30 cycles at 60Hz
Guidance and manufacturer’s declaration electromagnetic immunity
The NanoPix1/NanoPix2 is intended for use in the electromagnetic environment specified
below. The customer or the user of the NanoPix1/NanoPix2 should assure that it is used in
such an environment.